Table 1. Certified Values and Uncertainty (℃)
Sample Number |
Testing atmosphere |
Certified value and Uncertainty |
Characteristic Melting Temperature (℃) |
|||
Deformation Temperature (DT) |
Softening Temperature (ST) |
Hemispheric Temperature (HT) |
Flowing Temperature (FT) |
|||
SDRM 040005 579# |
Reducing |
Certified value |
1171 |
1184 |
1192 |
1217 |
Uncertainty |
21 |
30 |
30 |
30 |
||
Oxidizing |
Certified value |
1191 |
1202 |
1215 |
1250 |
|
Uncertainty |
26 |
30 |
33 |
15 |
||
SDRM 040006 580# |
Reducing |
Certified value |
1350 |
1366 |
1373 |
1389 |
Uncertainty |
15 |
15 |
17 |
18 |
||
Oxidizing |
Certified value |
1368 |
1380 |
1387 |
1419 |
|
Uncertainty |
21 |
23 |
24 |
26 |
||
SDRM 040007 581# |
Reducing |
Certified value |
1245 |
1294 |
1314 |
1374 |
Uncertainty |
18 |
15 |
15 |
29 |
||
Oxidizing |
Certified value |
1308 |
1344 |
1369 |
1409 |
|
Uncertainty |
23 |
20 |
17 |
15 |
The reducing atmosphere is obtained by introducing into the furnace the mixture gases of (50±5)% CO2 and (50±5)% H2 (in the most tests) or by sealing in the furnace the proper ratio of graphite and anthracite (in a few tests); the oxidizing atmosphere is obtained with air circulates freely through the furnace.
The certification analyses were performed according to the Chinese National Standard GB/T219-2008 by 8 qualified laboratories.
The homogeneity tests were performed according to GB/T219-2008, the results show that the homogeneity was good.
Expiration of certification: The certification of this CRM is valid until November 1, 2031.
The certified reference material is packed in glass bottles with plastic covers. The net weight is 25g each. It is suggested to keep dryness when stored. The certified reference material should be reduced in size to <0.1mm and ignited at 815℃ for 1 hour before use. This certified reference material is not applicable in the identification of the deviation of furnace temperature, the users should be sure the furnace temperature has been controlled correctly before the testing.
Product detail pictures:
Related Product Guide:
We not only will try our greatest to offer you excellent services to just about every client, but also are ready to receive any suggestion offered by our buyers for Wholesale Crm Excel Suppliers – Reference Material For Steel Testing – Shandong Institute of Metallurgical Science , The product will supply to all over the world, such as: Sydney, Juventus, Romania, Now we've been sincerely consider to grant brand agent in different areas and our agents' maximum margin of profit is the most important thing we care about. Welcome all of the friends and customers to join us. We have been ready to share win-win corporation.