Reference Material For Steel Testing

products

Reference Material For Steel Testing

Short Description:

This certified reference material can be used for checking correctness of testing atmosphere in ash fusibility determination. It can also be used in quality control of the analysis process and method evaluation.


Product Detail

Product Tags

Product introduction

Steel Testing (1)
Ash fusibility (2)

Certified Values

Table 1. Certified Values and Uncertainty (℃)

Sample Number

Testing atmosphere

Certified value and Uncertainty

Characteristic Melting Temperature (℃)

Deformation Temperature

(DT)

Softening

Temperature

(ST)

Hemispheric

Temperature

(HT)

Flowing

Temperature

(FT)

SDRM

040005

579#

Reducing

Certified value

1171

1184

1192

1217

Uncertainty

21

30

30

30

Oxidizing

Certified value

1191

1202

1215

1250

Uncertainty

26

30

33

15

SDRM

040006

580#

Reducing

Certified value

1350

1366

1373

1389

Uncertainty

15

15

17

18

Oxidizing

Certified value

1368

1380

1387

1419

Uncertainty

21

23

24

26

SDRM

040007

581#

Reducing

Certified value

1245

1294

1314

1374

Uncertainty

18

15

15

29

Oxidizing

Certified value

1308

1344

1369

1409

Uncertainty

23

20

17

15

The reducing atmosphere is obtained by introducing into the furnace the mixture gases of (50±5)% CO2 and (50±5)% H2 (in the most tests) or by sealing in the furnace the proper ratio of graphite and anthracite (in a few tests); the oxidizing atmosphere is obtained with air circulates freely through the furnace.

Analysis Methods

The certification analyses were performed according to the Chinese National Standard GB/T219-2008 by 8 qualified laboratories. 

Homogeneity Test and Stability Inspection

The homogeneity tests were performed according to GB/T219-2008, the results show that the homogeneity was good.

Expiration of certification: The certification of this CRM is valid until November 1, 2031. 

Packing and Storage

The certified reference material is packed in glass bottles with plastic covers. The net weight is 25g each. It is suggested to keep dryness when stored. The certified reference material should be reduced in size to <0.1mm and ignited at 815℃ for 1 hour before use. This certified reference material is not applicable in the identification of the deviation of furnace temperature, the users should be sure the furnace temperature has been controlled correctly before the testing.

Laboratory

Name: Shandong Institute of Metallurgical Science Co., Ltd.

Address : 66 Jiefang East Road, Jinan, Shandong, China;

Website: www.cncrms.com

Emai: cassyb@126.com

New standard coal1

Approved by: Gao Hongji

Laboratory Director

Date: December 1, 2020


  • Previous:
  • Next:

  • Write your message here and send it to us