Table 1. Certified Values for YSBC 28604a-2013 (Mass Fraction %)
Number |
Elements |
C |
S |
Si |
Mn |
P |
Fe |
Al |
YSBC 28604a-2013 |
Certified Values |
0.050 |
0.002 |
69.96 |
0.256 |
0.023 |
26.57 |
1.26 |
Uncertainty |
0.003 |
0.001 |
0.10 |
0.004 |
0.001 |
0.06 |
0.02 |
|
Number |
Elements |
Ca |
Cu |
Ni |
Cr |
Ti |
Mg | |
YSBC 28604a-2013 |
Certified Values |
0.649 |
0.013 |
0.010 |
0.049 |
0.107 |
0.019 |
|
Uncertainty |
0.005 |
0.001 |
0.001 |
0.002 |
0.004 |
0.001 |
Table 2. Analysis Methods
Composition |
Method |
C |
The Infrared absorption method
Gas-volumetric method |
S |
The Infrared absorption method
Combustion iodometric method |
Si |
Perchloric acid dehydration gravimetric method
Silicon potassium fluoride titrimetric method |
Mn |
Periodate photometric method
ICP-AES |
P |
Bismuth phosphomolybdate blue spectrophotometric method
Antimony phosphomolybdate blue spectrophotometric method ICP-AES |
Fe |
Potassium dichromate titration method |
Al |
EDTA titrimetric method
The chrome azurol S photometric method ICP-AES |
Ca |
ICP-AES
AAS |
Cu |
BCO spectrophotometric method
ICP-AES AAS |
Ni |
Dimetylglyoxime spectrophotometric method
ICP-AES |
Cr |
Diphenyl carbon acyl dihydrazide photometry
ICP-AES |
Ti |
Diantipyryl methane photometric method
ICP-AES |
Mg |
ICP-AES
AAS |
Expiration of certification: The certification of this CRM is valid until November 1, 2023.
Table 3. Methods for homogeneity testing
Composition |
Analysis methods |
Minimum sample (g) |
Si |
Silicon potassium fluoride titrimetric method |
0.1 |
C、S |
The Infrared absorption method |
0.2 |
Mn、P、Al、Ca、Cu、Ni、Cr、Ti、Mg |
ICP-AES |
0.2 |
Fe |
Potassium dichromate titration method |
0.2 |
The certified reference material is packed in glass bottles with plastic covers. The net weight is 50 g each. It is suggested to keep dryness when stored.
Product detail pictures:
Related Product Guide:
Our well-equipped facilities and excellent excellent management throughout all stages of creation enables us to guarantee total buyer satisfaction for China In House Reference Standard Manufacturer – Reference Material For Instrument Calibration – Shandong Institute of Metallurgical Science , The product will supply to all over the world, such as: Durban, San Francisco, Belize, To let customers be more confident in us and get the most comfortable service, we run our company with honesty, sincerity and best quality . We firmly believe that it is our pleasure to help customers to run their business more successfully, and that our professional advice and service can lead to more suitable choice for the customers.
By Cara from Sydney - 2018.12.14 15:26
We are long-term partners, there is no disappointment every time, we hope to maintain this friendship later!
By Margaret from Japan - 2018.09.12 17:18