Table 1. Certified Values for YSBC 15602-2006 (Mass Fraction %)
Number |
Elements |
Ti |
C |
Si |
Mn |
P |
S |
Cr |
YSBC 15602-2006 |
Certified Values |
70.02 |
0.057 |
1.47 |
0.106 |
0.0071 |
0.0047 |
0.039 |
Uncertainty |
0.17 |
0.003 |
0.03 |
0.004 |
0.0005 |
0.0006 |
0.003 |
|
Number |
Elements |
Ni |
Mo |
V |
Cu |
Al |
Fe |
|
YSBC 15602-2006 |
Certified Values |
0.29 |
0.028 |
0.011 |
0.037 |
0.30 |
26.57 |
|
Uncertainty |
0.02 |
0.003 |
0.002 |
0.003 |
0.02 |
0.13 |
Table 2. Analysis Methods
Composition |
Method |
Ti |
The Ammonium ferric sulfate titration method |
C |
The Infrared absorption method |
Si |
ICP-AES Perchloric acid dehydration gravimetric method The silicomolybdic blue spectrophotometric method The Sulphuric acid dehydration gravimetric method |
Mn |
Periodate photometric method AAS ICP-AES |
P |
Molybdobismuthylphosphoric blue spectrophotometric method Molybdenum blue spectrophotometric method Molybdenum blue photometry for extraction by butyl acetate Extraction of Molybdenum Blue from N-Butyl and Alcohol-Third Chloromethane ICP-AES |
S |
The Infrared absorption method |
Cr |
ICP-AES AAS Diphenyl carbon acyl dihydrazide photometry |
Ni |
ICP-AES AAS Budione oxime photometry |
Mo |
Rhodanate-butyl acetate extraction photometryd ICP-AES |
V |
Tantalum reagent extraction spectrophotometric method ICP-AES |
Cu |
ICP-AES AAS The DDTC photometric method |
Al |
ICP-AES EDTA titrimetric method The chrome azurol S photometric method |
Fe |
ICP-AES Dichromate titration method |
Expiration of certification: The certification of this CRM is valid until January 1, 2026.
Table 3. Methods for homogeneity testing
Composition |
Analysis methods |
Minimum sample (g) |
Ti |
The Ammonium ferric sulfate titration method |
0.2 |
C、S |
The Infrared absorption method |
0.2 |
Mn、P、Cr、Ni、Mo、V、Cu、Al |
ICP-AES |
0.1 |
Si |
Perchloric acid dehydration gravimetric method |
0.5 |
Fe |
Dichromate titration method |
0.2 |
The certified reference material is packed in glass bottles with plastic covers. The net weight is 50g each. It is suggested to keep dryness when stored.
Product detail pictures:
Related Product Guide:
Our goal is to satisfy our customers by offering golden service, good price and high quality for China Analytical Reference Materials – Reference Material For Environmental Testing – Shandong Institute of Metallurgical Science , The product will supply to all over the world, such as: New Orleans, United States, Tunisia, With the goal of "zero defect". To care for the environment, and social returns, care employee social responsibility as own duty. We welcome friends from all over the world to visit and guide us so that we can achieve the win-win goal together.